Service manual

Bair Hugger Model 775 Temperature Management Unit Service Manual
16
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Viewing the Fault Code Log
Service frequency
As oen as necessary
Tools/equipment
None required
Method
1� If your temperature management unit is already ON, you can access the fault code log from any
operating mode� Otherwise, connect the temperature management unit to a grounded power source�
e unit performs a self-test and automatically proceeds to Standby mode� en select any operating
mode
2� Press the Alternative Mode button ("Figure 2� Location of Alternative Mode button" on page11)
and the 32°C button simultaneously for three seconds� e temperature management unit is now in
Fault Code Log mode� In this mode, the temperature management unit performs the following tasks:
a� Continues to operate in current operating mode�
b. 32°C and 38°C indicator light ashes
c� Text “FC(n) XXX” appears in the alphanumeric display� “FC” is an abbreviation for the
phrase “fault code,” “n” represents the numerical order of the fault code in the log, and “XXX
represents the code for the fault condition� See page 17 for a list of fault codes and their
meanings
d� e time (HH:MM:SS) of each fault appears below “FC(n) XXX” in the alphanumeric display� If
the unit reaches 100 or more hours the time only reads hours and minutes
3� Press either the 32°C or 38°C button to view the next fault code in the log� e temperature
management unit stores a maximum of 5 fault codes in non volatile memory
4� Press the Alternative Mode button to return the temperature management unit to its previous
operating mode�
Clearing the Fault Code Log
Put the unit in Fault Code mode by pressing the Alternative Mode button and the 32°C button
simultaneously for three seconds; then, press the 43°C button�
note
e fault code log should not be cleared unless instructed to do so by a 3M Patient Warming representative�