Service manual

Bair Hugger Model 775 Temperature Management Unit Service Manual
22
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
8� Be sure power and heater wires are routed away from each other
9� Reconnect the UI wire and slide the upper enclosure over the chassis
10 Reinstall the four enclosure screws to attach the upper enclosure to the back plate� Push down on the
top of the unit when tightening the screws to attain proper sealing of the enclosures�
11� Flip the unit upside down and replace the eight enclosure screws� In order to obtain a proper seal
and prevent noise, be sure to tighten these screws fully� If using a power screwdriver, you may need
to increase the torque setting to ensure proper seating� Once complete, if there is a whistling noise
present it is most likely from these screws not being fully tightened�
12 Reattach the hose (see Reattaching Hose on page 20 for instructions)�
General Maintenance and Storage
Cleaning the Warming Unit
Service frequency
As needed�
Tools/equipment
• So cloth lightly dampened with water
• Mild detergent or antimicrobial spray
• Dry so cloth
warning
• Do not immerse the cabinet or hose while cleaning� Moisture will damage the components, and
thermal injury may result�
cautions
• Do not use a dripping wet cloth to clean the cabinet� Moisture may seep into the electrical contacts
and damage the components�
• Do not use alcohol or other solvents to clean the cabinet� Solvents may damage the labels and other
plastic parts�
Method
1� Disconnect the temperature management unit from the power source before cleaning�
2� Wipe the cabinet and the outside of the hose with a damp, so cloth and a mild detergent or
antimicrobial spray
3� Dry with a separate so cloth
Storage
Store all components in a cool, dry place when not in use