Bair Hugger Model 775 Temperature Management Unit Service Manual
English 202526A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
Low fan setting
Nonexplosion proof
Consult instructions for use
Follow instructions for use
Recycle to avoid environmental contamination� is product contains recyclable parts� For
information on recycling - please contact your nearest 3M Service Center for advice�
Date of manufacture
Manufacturer
Attention; see instructions for use
Keep dry
Temperature limit
Explanation of Signal Word Consequences
warning:
Indicates a hazardous situation which, if not avoided, could result in death or serious injury
caution:
Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury
notice:
Indicates a situation which, if not avoided, could result in property damage only
Indications
e Bair Hugger system is intended to prevent and treat hypothermia� In addition, the Model 775 Temperature
Management System can be used to provide patient thermal comfort when conditions exist that may cause
patients to become too warm or too cold� e temperature management system can be used with adult and
pediatric patients
Contraindications
Do not apply heat to lower extremities during aortic cross-clamping� ermal injury may occur if heat is
applied to ischemic limbs�
WARNINGS
1� Do not leave patients with poor perfusion unmonitored during prolonged warming therapy sessions�
ermal injury may result�