Bair Hugger Model 775 Temperature Management Unit Service Manual
4
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
2� e Bair Hugger temperature management unit has been designed to operate safely ONLY with 3M
Patient Warming disposable components� Use with other products may cause thermal injury� (To
the full extent permitted by law, the manufacturer and/or importer declines all responsibility for
thermal injury resulting from the unit being used in conjunction with products other than 3M Patient
Warming products�)
3� Do not warm patients with the temperature management unit hose alone� ermal injury
may result� Always connect the hose to a Bair Hugger blanket or Bair Paws gown before
providing therapy
4� Do not place the non-perforated side of the blanket on the patient� ermal injury may result� Always
place the perforated side (the side with small holes) towards the patient�
5� Do not continue temperature management therapy if the Over-temp indicator light illuminates
and the alarm sounds� ermal injury may result� Unplug the unit, and contact a qualied service
technician�
6� Do not continue 241 blood/uid warming therapy if the Over-temp indicator light illuminates and
the alarm sounds� Immediately stop uid ow, and discard the blood/uid warming set� Unplug the
temperature management unit, and contact a qualied service technician
7� Do not use a forced-air warming device over transdermal medication� Increased drug delivery and
patient injury or death may occur�
8� Do not allow the patient to lie on the warming unit hose or allow the hose to directly contact the
patient’s skin during patient warming; thermal injury may result�
9� Reusable blankets made from woven fabric, or blankets without discrete, visible holes, can cause the
safety system of this unit to fail, which may result in serious thermal injury� is warming unit has
been designed to operate safely ONLY with Bair Hugger blankets and Bair Paws gowns
10 Do not connect a Bair Hugger blanket, 241 blood/uid warmer, or Bair Paws gown to the warming
unit if it has been cut or damaged, thermal injury may result
11� Do not use a Bair Hugger blanket to transfer or move the patient; injury may result�
12 To reduce the risks associated with hazardous voltage and re:
• keep power cord visible and accessible at all times� e plug on the power cord serves as the
disconnect device� e wall socket outlet shall be as close as practical and shall be easily accessible�
• use only the power cord specied for this product and certied for the country of use�
• do not allow the power cord to get wet
• do not use the warming unit when it appears the warming unit, power cord or any component is
damaged� Contact 3M Patient Warming Technical Support at 1-800-733-7775�
• this equipment must only be connected to a supply mains with protective earth
13� To reduce the risks associated with exposure to biohazards always perform the decontamination
procedure prior to returning the warmer for service and prior to disposal
14 Do not retain the patient using a warming blanket alone, as injury may result� Use a draw, sheet safety
strap, or other means to retain the patient�
15� Do not modify this equipment without authorization of the manufacturer
16 To ground this Bair Hugger warming unit, only connect to receptacles marked “Hospital Only,
“Hospital Grade,” or a reliable grounded outlet