User Manual

20
Memory mode
You can store measurements on the tester:
• Telaris ProInstall-100 – up to 399
• Telaris ProInstall-200 – up to 1399
The information stored for each measurement consists of the test function and all user
selectable test conditions.
Data for each measurement is assigned a data set number, data subset number, and a data
id number. Memory location fields are used as described below.
Field Description
Use the data set field (a) to indicate a location such as a room or
electrical panel number.
Use the data subset field (b) for circuit number.
The data id number field (c) is the measurement number. The
measurement number automatically increments. The measurement
number can also be set to a previously used value to overwrite an
existing measurement.

1. Press the
to enter Memory mode.
The display changes to a memory mode display. In Memory mode, the
icon
appears on the display.
The primary numeric display shows the data set number (a, 1-9999). The secondary
numeric display shows the data subset number (b, 1-9999). The data id number (c,
1-9999) appears after you press F1 several times. One of the memory locations, a, b, or c,
will flash to indicate that you can change the number using the arrow keys
.
2. To enable the data subset number to be changed, press F1. The data subset number will
now be flashing. To enable the data sub number to be changed, press F1 again. The data
set number will now be flashing. Press F1 again to change the data id number.
3. Press the down arrow key (
) to decrement the enabled number or press the up arrow
key (
) to increment the enabled number. For storing data, the number can be set to
any value, overwriting existing data is allowed. For recalling data, the number can only
be set to used values.
Note: If you press the up or down arrow key (
) once, the number increments or
decrements by one. To accelerate the increment or decrement function, press and hold
the up or down arrow.


1. Press
to enter Memory mode.
2. Press F1 and use the arrow keys (
) to set the data identity
3. Press F2 to save the data.
•If memory is full, FULL will appear on the primary display. Press F1 to choose another
data identity, press
to exit Memory mode.