Manual

C1000 Touch
TM
Thermal Cycler Manual
18
Lid temperature determines the temperature of the heated lid. If the temperature
is too high, the sample temperature might rise above the target temperature
The C1000 Touch thermal cycler provides two ways of managing and entering sample volume
and lid temperature settings:
Change the setting when creating and editing the protocol (page 18)
Change the setting while initiating a run (page 21)
Temperature Control Modes
The C1000 Touch thermal cycler uses one of two temperature control modes to determine
when the sample reaches the target temperature:
Calculated mode When a sample volume of greater than zero is entered, the
thermal cycler calculates the sample temperature based on the sample volume. The
calculated mode is recommended, because it most accurately represents the actual
sample temperature
Block mode — when a sample volume of zero (0) μl is entered, the thermal cycler
assumes that the sample temperature is the same as the measured block
temperature
Choosing Lid Temperature
The adjustable heated lid of the C1000 Touch thermal cycler allows the user to control the lid
temperature and force. Heating the lid prevents condensation from forming inside the reaction
wells. When the C1000 Touch thermal cycler is running, the heated lid maintains the
temperature specified for the protocol being run. Without a heated lid, water can be lost from
the reagents to condensation, concentrating the reactants in the tube or plate.
The default lid temperature of the C1000 Touch thermal cycler is 105°C for 96- or dual 48-well
reaction blocks and 95°C for 384-well blocks.
NOTE: When the block is running an infinite hold at a temperature below 30.0°C,
the lid heater maintains 31.0°C.
Changing Sample Volume
To change the sample volume from the new protocol or protocol editor screen:
1. Touch the Volume button.
2. Enter a value using the pop-up numeric keypad.
3. Touch the OK button to confirm.
NOTE: Entering a value greater than zero in the Volume button selects Calculated
Temperature control mode, which is the standard mode. Entering zero (0) in the
Volume button selects Block mode. Calculated mode is the recommended mode
because it most accurately represents the actual sample temperature.
Changing Lid Temperature
When using a 96- or dual 48-well reaction block, the recommended lid temperature is 105°C.
When using a 384-well block, the recommended lid temperature is 95°C.
To change the lid temperature from the new protocol or protocol editor screen:
1. Touch the Lid button.