Manual

C1000 Touch
TM
Thermal Cycler Manual
22
In Figure 18, the protocol will start running on both blocks at the same time.
Figure 18. Multiple blocks running the same protocol at the same time.
5. Confirm or change the run parameters before starting the run.
TIP: The value entered in the Sample Volume field determines the temperature
control mode. If a sample value greater than 0 (zero) is entered, the instrument uses
Calculated Mode to calculate the target temperature. If 0 (zero) is entered in the
Sample Volume field, the Block Mode is used, in which case the thermal cycler
assumes that the temperature of the sample is the same as that of the block.
In Figure 19, the run parameters include the sample volume of 10ul and the lid
temperature of 105°C. (Touch the OK button to bypass these fields and run the protocol
without entering additional parameters).
Figure 19. The run parameters of the protocol.
6. To add optional additional settings, touch the Additional Settings button to enter a user
name, email address (to send an email notification upon run completion), sample ID or to
set a ramp rate for all steps of a protocol (Figure 20). Also, run a protocol on the C1000
Touch thermal cycler at the same ramp speed as that of the DNA Engine® thermal cycler
by touching the DNA Engine Ramp checkbox.