Manual

C1000 Touch
TM
Thermal Cycler Manual
25
2. To resume the protocol, touch the Resume button.
Skipping Steps in a Protocol
Skip a step if it is necessary to shorten a protocol while it is still running. By repeatedly
skipping steps, it is possible to bypass several cycles of a GOTO loop and shorten a protocol.
1. To skip the current protocol step in the status screen, touch the Skip button.
2. Touch the Skip button multiple times to skip more than one step.
Canceling a Run
A protocol may be canceled while it is running. When the protocol is canceled, the block
immediately stops changing temperature.
NOTE: Do not turn the thermal cycler off immediately after canceling a run. The
fans should be allowed to run long enough to cool the block.
To cancel a protocol running on a single block:
1. Open the status screen.
2. Touch the Cancel button to stop the protocol running on a single block.
NOTE: To cancel a protocol running on multiple blocks, select each block
individually from the dropdown list and touch the Cancel button. Continue until the
protocol is canceled on all blocks.
Incubating Samples
You may keep samples at a constant temperature for any amount of time.
WARNING! Incubating samples for extended periods of time at 4–10°C,
particularly in areas of high humidity, can cause excessive moisture condensation
around the block.
To start incubating samples at a single temperature:
1. Touch the INCUBATE button on the home screen.