4.1 Specifications
Weight 56.6 lb. (26 kg)
Size (W x D x H) 26 x 20.5 x 11.5" (660 x 521 x 292 cm)
Electrical requirements 100–240 V, 50/60 Hz, 1 A; voltage fluctuations
not to exceed +10% of ratings (for external power
supply provided)
Temperature 18–30ºC
Altitude 06,560 ft (0–2,000 m)
Humidity 85% max (noncondensing)
Pollution degree 2 (indoor use)
Installation category II (external power supply plugs into standard AC
receptacle)
Ventilation requirements 5" (12 cm) left and right of machine and 10" (25 cm)
behind should be unobstructed for proper ventilation
11. 5 in ch es
(292 mm)
26 inches
(660 mm)
20.5 inches
(521 mm)