Full Product Manual
Product Descripon
The bite away
®
is a patented handheld device that allows the user to apply a therapeuc dose of heat directly to an insect sng
or bite and is clinically proven to provide temporary relief of the resulng itching, pain and swelling. If the device is applied
immediately aer the sng or bite, the symptoms can usually be prevented completely. Even if the device is used slightly later,
the symptoms can sll subside sooner.
bite away
®
works solely by concentrated heat and is thus completely free from chemicals (also suitable for pregnant women,
allergy suerers and children).
The bite away
®
allows the user to choose either a 3 second heat dose or a longer 5 second heat dose. The ceramic contact
surface is heated up to 122 - 127°F (50 - 53°C) by an integrated micro-processor and maintained for 3 or 5 seconds according
to the buon selected.
The bite away
®
is provided with two AA LR6 1.5V baeries to power the device, which have to be placed into the baery
compartment before the rst use (see illustraon below).
Indicaons for Use
The bite away
®
is indicated for use to provide temporary relief of the itching and pain resulng from insect sngs and bites such
as bees, wasps and mosquitoes by temporarily increasing localized blood ow.
User group/paent group:
adults and children aged 12 years and older in self-applicaon / under 12 years only in third party applicaon by adult person
Environment:
all environments except oxygen rich environment
Contraindicaons
– Not for use on mucus membranes (e.g. inside of the mouth, nose, etc.).
– Not for use on or near the eyes. Do not simultaneously use with medicaon and/or cosmecs.
– Not for use in case of heavy or chronic skin inammaon or severe internal diseases, in which the use could cause deterio-
raon, such as:
• open wounds, damaged skin
• scalded or burnt skin
• frostbite
• small-ber neuropathy, allodynia, hyperalgesia
• tumors and malignant changes of the skin
• people with reduced sensivity due to physical or medical complaints.
Direcons of Use
– Remove the snger before applying the bite away
®
to the site of the sng.
– Place the contact surface (round ceramic disc at the p of the device) directly onto the skin where the bite or sng had
occurred and gently press it against the skin.
– Choose either:
3 second treatment me (le buon) for the inial treatment and sensive skin (e.g. paents under 12 years of age in
third party applicaon) or
5 second treatment me (right buon) for regular applicaon.
– Press the buon and release immediately. The ceramic contact surface is heated up to a temperature of around 124°F
(51°C). Once this temperature is reached, the LED will light up. The LED will go o once the device automacally shuts
down aer the end of the chosen treatment me. At the start of the heang-up phase and at the end of the treatment, you
will hear an addional acousc signal.
– For successful treatment, do not interrupt the indicated duraon of the treatment. The treatment of the same insect sng
or bite may be repeated as required. Please leave a treatment interval of at least 2 minutes between each applicaon. Do
not exceed a maximum number of 5 applicaons per hour on the same insect sng or bite.
– The device packaging includes two baeries (2x AA LR6 1.5V). Empty baeries of the bite away
®
are indicated when LED
ashes 3 mes and triple acousc signal is heard.
!
Warnings
– Use carefully - may cause serious burns. Do not use over sensive skin areas or in the presence of poor circulaon. The
unaended use of the bite away
®
by children or incapacitated persons may be dangerous.
– To avoid contaminaon to other person, the bite away
®
is intended to be used by only one person.
Do not apply one bite away
®
to more than one person.
– The heated ceramic surface may reach a maximum temperature of 127°F (53°C).
– If you may use the device repeatedly on the same sng or bite, insure a pause of at least 2 minutes before repeated applica-
on. Do not exceed a maximum number of 5 applicaons per hour on the same insect sng or bite.
– The bite away
®
is not intended to replace medical advice or treatment. If symptoms (e.g. pain) are severe or persistent,
consult a physician.
In the event of severe hypersensivity reacons aer an insect sng or bite (anaphylacc reacon), bite away
®
is not a re-
placement for emergency treatment. You should immediately call a doctor at the rst signs of any such reacon (e.g. increased
itching, reddening, swelling or formaon of a welt, dizziness, nausea, dicules breathing, hypotension or hypertension).
Medical advice should be sought in the following cases:
• persistent symptoms while using bite away
®
.
• worsening symptoms while using bite away
®
. Unl these symptoms have been claried, you should stop using the device.
– Do not use on or near the eye area or on mucus membranes (e.g. inside the mouth or nose).
– Individuals with impaired sensivies due to physical or medical condions should use extreme cauon when using the bite
away
®
. If in queson, consult a physician prior to use.
– Do not use the bite away
®
if damaged or in case of obvious malfuncon. Do not use damaged baeries.
– For external use only.
– The use of other accessories as listed in this IFU may negavely aect the EMC performance or lead to noncompliance.
– If you intend to not use the bite away
®
for a longer me, please remove the baeries to prevent them to leak out.
– The bite away
®
is not intended to be used in oxygen rich environment.
– Please keep away the bite away
®
from young children or babies, because small parts (like a disassembled baery cap) could
be swallowed.
– Electromagnec interference (EMI) from common household products such as security systems, metal detectors, diathermy
equipment, MRI devices, RFID (Radio Frequency Idencaon) eming devices may interfere with the bite away
®
device.
The bite away
®
should not be used adjacent to or stacked with other electromagnec equipment. If the bite away
®
device
is adjacent or stacked with other electromagnec equipment , assure that the bite away
®
operaon is normal by verifying
that the LED light illuminates and the device emits a ‘beep’ when the le and or right buon is pressed prior to contacng
the device with the skin. You can help prevent electromagnec interference by maintaining a minimum distance between
portable and mobile RF communicaons equipment (transmiers) and the bite away
®
.
– Contact with liquids: Do not soak in water, wash or lubricate the bite away
®
, this may damage it. Do not use water, chemicals
or other cleaning soluons to clean the bite away
®
. Furthermore, the device must not be used on moist skin.
!
Precauons
– Not recommended for children under the age of 2 years.
– The bite away
®
is indicated only for self-administraon by adults and children aged 12 years and older. Children under the
age of 12 years should only use the device under the supervision of an adult.
– The bite away
®
is intended for the relief of itching and pain immediately aer an insect sng or bite. In case of an allergy to
these insects (or in the case of connuous pain), consult a physician immediately.
– The bite away
®
is not eecve against venom from spiders, scorpions or snakes.
– The use of bite away
®
may cause reddening of the skin.
– The use of bite away
®
may cause a burn blister if used on the same skin area more than once within a short period of me.
– The use of bite away
®
may cause a snging sensaon or a transient hot feeling.
Maintenance
A periodic maintenance is not dened. Please clean the bite away
®
or replace the baeries if necessary.
Cleaning and Storage
– Please clean the outside of bite away
®
by wiping it with a clean cloth.
– Do not soak in water, wash or lubricate the bite away
®
, this may damage it.
– Do not use water, chemicals or other cleaning soluons to clean the bite away
®
.
– Clean o dirt and dust with a dry cloth.
– Please store the bite away
®
on a safe place, dry, clean, no extreme heat and without direct sunlight.
– Please do not drop the bite away
®
on hard surface.
Warranty
2 years (valid from date of purchase)
Technical data
Heang mes
inial treatment and
persons with sensive skin
child (12 – 18 years of age)* or
child (under 12 years of age)** or
adult (over 18 years of age)*
regular applicaon
child (12 – 18 years of age)* or
child (under 12 years of age)** or
adult (over 18 years of age)*
3s
5s
Range of ceramic p temperature
target treatment temperature
122°F (50°C) – 127°F (53°C)
124°F (51°C)
Internal voltage 2.4 + 3.0 V (2x AA LR6 1.5V baeries)
shelf life of baery: please refer to baery labeling
! Only original baeries containing labeling for shelf life are allowed for use with
the bite away
®
!
* self-applicaon
** only in third party applicaon by adult
Accessories
– 2x AA LR6 1.5V baeries (no rechargeable baeries allowed)
The use of other accessories as listed in this IFU may negavely aect the EMC performance or lead to noncompliance!
Removal or Replacement of Baeries
2x AA LR6 1.5V baeries included in delivery. Before the rst use, place the baeries
into the baery compartment as shown in the illustraon.
Please remove the baeries when you intend to not use the bite away
®
for a longer me,
to prevent the baeries to leak out.
A storage of the baeries at cool temperatures (<30°C) will obtain their lifeme.
If the bite away
®
indicates empty baeries, please replace them.
Environmental Condions
The bite away
®
is intended for storage and operaon in a room-temperature environment.
Condion Temperature Humidity
Pressure
(Altude)
Operang 08°C – 40°C 15% – 93% 700 hPa – 1060 hPa
Storage 08°C – 40°C* 15% – 93% 700 hPa – 1060 hPa
Transport 08°C – 40°C 15% – 93% 700 hPa – 1060 hPa
* in case of longer storage without baeries (please remove the baeries and store them cool < 30°C)
Symbols
Refer to manual
!
Cauon:
Consult instrucons for use
Type BF Applied Part CLASS II equipment
Batch Code
or
Manufacturer
12
6
3
9
1
2
4
5
7
8
10
11
Charge clock Protect against UV rays
Operang temperature
Storage temperature
or
Protect against liquids
Keep away from rain
IP42
Protecon against Ingress
of Solids and Liquids
Disposal of waste products
%
Humidity condions
MR
MRI Unsafe – Keep away
from magnec resonance
imaging (MRI) equipment
Instrucons for Use
PRESS
1
OPEN
2
INSERT BATTERIES
3
86 °F
47 °F
GI03322-01_bite_away_US.indd 1 03.05.2019 11:16:17