User guide

8818 User Guide
(BB1505-F) Puncture and Biopsy Facilities
13
Endfire
The sterile single-use biopsy guide UA1323-S and reusable biopsy guide UA1327
are used in endfire imaging.
Figure 7. Reusable biopsy guide UA1327.
The puncture line for UA1323-S and UA1327 on 8818 is shown in Figure 8. The
needle guide is parallel to the centerline of the transducer.
Figure 8. Illustration of the puncture line for biopsy guides UA1323-S and UA1327.
Dual
The sterile single-use biopsy guide UA1329-S and reusable biopsy guide UA1328
are used in dual imaging. Needle guide UC0100 is marked with a blue band (see
Figure 9). On screen, you will see this corresponds to the color of the puncture line
displayed on the image.
Figure 9. Reusable biopsy guide UA1328.
WARNING
Before beginning a puncture or biopsy procedure using dual imaging, ensure that the color
of the puncture line on the system monitor matches that of the needle guide you will insert
your needle into.