User manual

114
Format:
ON or OFF
PROTection:OVP:LEVel
Description:
Configures/Queries the voltage level for OVP.
Syntax:
PROTection:OVP:LEVel<SP><voltage>
PROTection:OVP:LEVel?
<voltage> - 0 max. rated voltage.
Example:
PROT:OVP:LEV 20.000
Set 20.000 V as the OVP voltage level.
Query:
PROT:OVP:LEV?
Return String:
20.000
Format:
<NRf>
PROTection:OVP2
Note:
For dual channel models only.
Description:
Configure/Query the state of OVP (over voltage
protection) for channel 2.
Syntax:
PROTection:OVP2<SP><Boolean>
PROT:OVP2?
Example:
PROT:OVP2 1
PROT:OVP2 ON
Enables OVP for channel 2.
Query:
PROT:OVP2?
Return String:
OFF
Currently OVP is disabled for channel 2.
Format:
ON or OFF
PROTection:OVP2:LEVel
Note:
For dual channel models only.
Description:
Configures/Queries the voltage level for OVP for
channel 2.
Syntax:
PROTection:OVP2:LEVel<SP><voltage>
PROTection:OVP2:LEVel?
<voltage> - 0 max. rated voltage.
Example:
PROT:OVP2:LEV 20.000
Set 20.000 V as the OVP voltage level for channel 2.
Query:
PROT:OVP2:LEV?
Return String:
20.000
Format:
<NRf>