User manual
156
Programming and Readback Accuracy ±(% output+offset)
Voltage
≤ 0.05%+5mV(9171,9173,9181, 9182)
≤ 0.05%+10mV(9172,9174,9183)
≤ 0.05%+50 mV(9184)
≤ 0.05%+100mV(9185)
Current
≤ 0.1%+2mA (9171,9173,9181,9183)
≤ 0.1%+1mA (9172,9174,9184)
≤ 0.1%+5 mA (9182)
≤ 0.1%+0.1 mA (9185)
Temperature Coefficient per °C ±(% output+offset)
Voltage
≤ 0.005%+1mV
≤ 0.005%+10 mV (9184)
≤ 0.005 %+20 mV (9185)
Current
≤ 0.01%+3mA
Stability (8 hrs) ±(% output+offset)
Voltage
≤ 0.02%+2mV
≤ 0.02 %+10 mV(9184)
≤ 0.02%+20 mV(9185)
Current
≤ 0.1%+1mA
Settling Time
2
≤ 30ms
Measurement Time
≤ 50ms
Transient Response Time
3
≤ 50μs (9171 - 9183)
≤ 100μs (9184, 9185)
OVP Accuracy
≤ 0.5%+0.1V (9171 – 9183)
≤ 0.5%+1V (9184, 9185)
OCP Accuracy
≤ 0.5%+0.1A
OVP/OCP Activation Time
4
≤ 1mS
Remote Sense
Compensation
1V Max
Rising Time at Full Load
≤ 8mS (9171,9173,9181,9182)
≤ 10mS (9172,9174,9183)