User manual

157
30mS (9184)
40mS (9185)
Rising Time at No Load
8mS (9171,9173,9181,9182)
10mS (9172,9174,9183)
30mS (9184)
40mS (9185)
Falling Time at Full Load
8mS (9171,9173,9181,9182)
10mS (9172,9174,9183)
30mS (9184)
40mS (9185)
Falling Time at No Load
250mS
Standard Interface
USB
Optional Interface Cards
5
LAN/GPIB Card, Digital I/O and Analog
Control Cards (1CH and 2CH), RS485
Card, RS232 Card
General
Power Requirements
115/230 VAC± 10%, 47Hz - 63Hz
Operating Temperature
0°C- 40°C
Storage Temperature
-10°C- 70°C
1
With sense terminal connected.
2
Maximum time required for the output voltage to change from 1% to 99% or
vice versa following the receipt of VOLTage or VSET command via direct
GPIB or USB interface.
3
Models 9171 9183: Less than 50 μs for output to recover to within 15 mV
following a change in output current from full load to half load or vice versa.
Model 9184: Less than 100 μs for output to recover to within 50 mV following
a change in output current from full load to half load or vice versa.
Model9185: Less than 100 μs for output to recover to within 120 mV following
a change in output current from full load to half load or vice versa.
4
Average time for output to start to drop after OVP/OCP condition occurs.
5
Digital I/O Analog Control Card Accuracy:±1%.
To ensure the most current version of this manual, please download the
latest version here: http://www.bkprecision.com/search/9171
For current up-to-date product information, please visit
www.bkprecision.com