User manual

162
IL Set = __.____A on the display. Press to proceed to IML.
If current shunt is used for the calibration, divide the DMM
readings (in DCV) by the resistance (Rs) of the shunt. The result
should be DCV/Rs. Enter this current result into IL.
4. The power supply will then pre-set another current value for IML.
Repeat the same process in step 3. Once finished, repeat the
same process for IMH and IH, as shown below.
5. Once finished, the OCP Lo
calibration will begin. Press
to start. After it’s complete, the unit will select CALIB OCP Hi for
calibration. Again, after it’s complete, press .
6. Once this calibration process is complete, the calibration data
will be stored into flash memory, and the supply will return to the
calibration menu.
IMH Set = X . XXXXA Read = XXXXX
IH Set = X . XXXXA Read = XXXXX
CALIB OCP Lo = Start
CALIB OCP Hi = Start
Enter
Enter
Enter