User manual

164
3. The power supply will output a pre-set voltage to the DMM. Wait
for the DMM reading to settle, then enter it into where it shows
VL Set = __.____V on the display. Press to proceed to the
next setp.
4. The power supply will then pre-set another voltage value for
VML. Repeat the same process in step 3. Once finished, repeat
the same process for VMH and VH, as shown below.
5. Once finished, the OVP Lo
calibration will begin. Press
to start. After it’s complete, the unit will select CALIB OVP Hi for
calibration. Again, after completion, press .
6. Once this calibration process is complete, the calibration data
will be stored into flash memory, and the supply will return to the
calibration menu.
VMH Set = XX . XXXXV Read = XXXXX
VH Set = XX . XXXXV Read = XXXXX
CALIB OVP Lo = Start
CALIB OVP Hi = Start
Enter
Enter
Enter