User manual

21
WARNING:
Do not connect power to the instrument until the
line voltage selection is setup correctly. Applying
an incorrect line voltage or configuring the line
voltage selection improperly may damage the
instrument and void all warranty.
2.3 Output Connections
These power supplies have both front panel binding posts and rear
panel terminals for output connections, and they are paralleled
together. The two (+) and two (-) (per channel for dual channel
models) rear output terminals can accept wire sizes AWG 24 to
AWG 12 (See Table 2
). However, we recommend using 12 AWG if
current output is between 5 to 10 A. For current output above 10 A,
remove the shorting barsfrom the front panel, but do not remove the
shorting pins on the rear panel.Use two separate wires/leads to
connect both (+) terminals and (-) terminals as shown below:
Line Voltage Switch
Rear P
anel
Front Panel
(Bottom View)
Front Feet
Rear Feet