Instructions

8
9
Error Description Potential cause and remedy
Readings from
repeated
measurements differ
considerably.
Blood pressure is a
fluctuating value.
For healthy adults,
deviations of 10 to
20 mmHg are possible.
Reminder: always use the same arm
for measurements!
Cuff is not properly
fitted.
Ensure that the cuff is fitted
according to the Applying the arm
cuff section.
Readings were not
consistently taken at
heart level.
For each measurement, make sure
that the cuff is at heart level.
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
While measuring, relax, keep still,
do not move or speak.
Blood pressure values
measured at your
doctors office differ
from your
measurements.
Doctor visits can
frequently cause
anxiety.
Record the daily development of
the measured values and consult
your doctor.
Display remains
blank, or is unusual,
when instrument is
switched on.
Batteries are not
installed properly.
Check batteries for correct polarity.
Device frequently
fails to measure
blood pressure
values, or values
measured are too low
(or too high).
Cuff may not be
properly positioned.
Ensure that the cuff is fitted
according to the Applying the arm
cuff section.
Further information
Blood pressure is subject to fluctuations even in healthy people.
Comparable measurements always require the same conditions
(quiet conditions)!
In order to receive market clearance from governmental bodies, this
device was subjected to strict clinical tests.
Specifications
Model: BUA5000
Operation principle: Oscillometric method
Display: Liquid crystal display
Cuff pressure display
range: 0-299 mmHg
Pulse measurement: 40-199 beats per minute
Blood pressure
measurement range: 30-280 mmHg
Laboratory accuracy: ±3 mmHg (cuff pressure)
±5% of reading (pulse rate)
Clinical accuracy: According to EN 1060-1/-3 and ISO 81060-2
with auscultatory reference:
<5 mmHg systematic offset
<8 mmHg standard deviation
Inflation: Automatic
Power supply: 4 batteries, type AA 1.5V
Operating temperature/
atmospheric pressure/
humidity: 10 °C to 40 °C (50 ˚F to 104 ˚F)
860 to1060 hPa (0.849 to 1.046 atm)
15% to 90% relative humidity, non-condensing
Storage temperature/
atmospheric pressure/
humidity: -20 °C to 55 °C (-1 ˚F to 131 ˚F)
860 to1060 hPa (0.849 to 1.046 atm)
15% to 90% relative humidity, non-condensing
Cuff: Universal cuff (22-42 cm)
Clinical validation: According to international protocol 2002 (ESH)
Service life: 5 years
Warranty: 2 years
If device is not used within specified temperature and humidity ranges the
technical accuracy cannot be ensured.
Type BF applied parts Operating temperature
See instructions for use
55
Storage temperature
Keep dry
Subject to change without notice.
Internally powered equipment.
Continuous operation.