Guidance and manufacturer’s declaration – electromagnetic
immunity
The BUA5000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BUA5000 should assure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kV Contact
±8kV Air
Complies
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Radiated RF
IEC 61000-4-3
3V/m 80MHz to
2.5GHz
Complies
Field strengths outside the
shielded location from fixed
RF transmitters, as determined
by an electromagnetic site
survey, should be less than
3V/m.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Separation distance
calculation provided above.
If a known transmitter is
present the specific distance
can be calculated using the
equations.
Conducted RF
IEC 61000-4-6
3Vrms 150kHz to
80MHz
Not
applicable
(no electrical
cabling)
Electrical fast
transient
IEC 61000-4-4
±2kV
power line
±1kV I/O lines
Not
applicable
The BUA5000
is solely battery powered.
Surge
IEC 61000-4-5
±1kV
differential
±2kV common
Not
applicable
Power frequency
magnetic field
IEC 61000-4-8
3A/m Complies
Power frequency
magnetic fields should be at
levels characteristic of a
typical location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
>95% dip 0.5 cycle
60% dip 5 cycles
70% dip 25 cycles
95% dip 5 sec.
Not
applicable
The BUA5000
is solely battery powered.