Product Description

8
9
Average function
Press the average button (5) to show last 3 readings average on LCD.
Erasing data
Switch o the device, press and hold memory button
M
(2) for
more than 5 seconds, LCD display will blink dEL ALL. Ensure
that the correct user; A or B, is selected before use.
Press memory button
M
(2) again, LCD will display “---” to show
all the stored data of corresponding user have been deleted.
Irregular heart beat detector
The appearance of this symbol signies that a certain pulse irregularity was detected during the
measurement. Talking, moving, shaking or an irregular pulse during the measurement can result in the
appearance of this icon. Usually this is not a cause for concern, however if the symbol appears often,
we recommend you seek medical advice. The device does not replace a cardiac examination, but
serves to detect pulse irregularities at an early stage.
Low battery indicator
When the low battery indicator ashes on the display, it means the battery is low and the four
batteries need to be replaced with alkaline LR6 (AA) batteries.
** After replacing batteries, ExactFit 3 will go into time-setting mode automatically and show
the last measurement time on screen. Please set the current date / time before taking the next
measurement to get correct average result.
Storage and cleaning
Always keep the unit in the carrying case after its use.
Do not put the item directly under the sunlight, in high temperature or humid and dusty places.
Do not store in extremely low (less than -20°C) or high (more than 60°C) temperature.
Use a piece of cloth with water or mild cleansing agent to clean the case and then use a piece of dry
cloth to wipe it dry. Use a piece of dry cloth to wipe the cu when it is dirty.
Do not use any strong cleansers to clean it.
When the unit is not to be used for a long time, remove the batteries.
(Batteries may leak or cause harm).
Do not modify the device. NEVER open the device! This will make the manufacturer’s warranty
invalid.
A
A
A
P
MD
A
What to do if …..
Problem Reason Solution
Heart rate
symbol
Appears in the measurement
condition and ashes when
pulse is detected.
Measurement in progress, remain quiet.
Low battery
indicator
Appears when the battery
voltage is excessively low or the
positions of the batteries are
incorrect.
Replace all four batteries with new ones.
Insert the batteries in the correct
positions. Be aware of +/- positions.
Measuring
error
Appears when the accurate
blood pressure and pulse could
not be obtained.
Press “start/stop button again and
remeasure.
Check if cu is wrapped according
to instructions.
Check that there are no kinks in tube.
Check that hand is relaxed.
Check if talking or moving
during measurement.
Check if posture is correct.
E1 shows The cu is not secure. Refasten the cu and then measure again.
E2 shows The cu is very tight. Refasten the cu and then measure again.
E 10 or
E 11 shows
The monitor detected motion,
talking or the pulse is too poor
while measuring.
Relax for a moment and then measure again.
E 20 shows The measurement process does
not detect the pulse signal.
Loosen the clothing on the arm and then
measure again.
E 21 shows Measure incorrectly. Relax for a moment and then measure again.
EE 3 - EE15 Error while measuring. Retake the measurement. If the problem
persists, contact the retailer or our customer
service department for further assistance.
Refer to the warranty for contact information
and return instructions.
English