User guide

- 13 -
9) If the program is to have less than 25 steps, the End Program Entry command is given by entering
“0” for the step speed as the final program step. If the program is to have 25 steps, entry is ended
automatically after the 25th step.
Next Mode Speed Set Notes:
a) The NEXT key has three functions:
1) Accept the value just entered.
2) Move the cursor between the speed and time field for the first step.
3) Move from step field to step field.
b) If any value other than zero (0) is entered as the time value in the first step, the program
will not be a Next Speed Set Program.
c) Zero (0) RPM is not a valid speed in Next mode.
III.2 Prog Mode Speed Set Program
To create a Prog Speed Set Program, proceed as follows:
1) Press the
PROG SPEED key.
2) Choose an available speed set program from the list on the screen using the numeric keys to enter
the location number. The screen, as explained in the preceding section, will automatically
switch from Figure 9 to Figure 10.
3) Enter the first step speed using the numeric keys. The speed must be between 0.0 rpm and 250.0
rpm, in increments of 0.1 rpm.
4) Press the NEXT key to accept the first step speed. The cursor moves to the time field.
5) Enter the first step time which may be any time other than 00:00 and must be between 1 second
and 99 minutes, 59 seconds. The format is MM:SS (Minutes Minutes:Seconds Seconds). The
default time is 00:05.
6) Press the NEXT key to accept the time entry. The cursor moves back to the speed field.
Note: If you wish to change any entry in either the speed or time field, you need only enter (write
over) the entry and accept again by pressing the NEXT key.
7) Press the PROG RUN key to accept the first step. The display changes to the second step.
8) Repeat steps 3 through 7 for as many steps (up to a maximum of 25 steps) as are required.
9) If the program is to have less than 25 steps, the End Program Entry command is given by entering
“0” for both step speed and time values as the final program step. If the program is to have 25
steps, the entry mode is ended automatically after the 25th step.
Prog Mode Speed Set Notes:
a) The NEXT key has two functions:
1) Accepts the value just entered.
2) Moves the cursor between the speed field and the time field.
b) In any Prog Speed mode program, the step time that you accept in any step is presented as the
default time for the next step.