User guide

- 40 -
Table D1 (continued)
Table D2 lists the model codes and spring torque constants for each Rheometer model.
The full scale viscosity range for any DV-III model and spindle may be calculated using the equation:
Where:
TK = DV-III Torque Constant from Table D2
SMC = Spindle Multiplier Constant listed in Table D1
SC4-27 27 25 0.34
SC4-28 28 50 0.28
SC4-29 29 100 0.25
SC4-31 31 32 0.34
SC4-34 34 64 0.28
SC4-37 37 25 0.36
CP40 40 0.327 7.5
CP41 41 1.228 2
CP42 42 0.64 3.8
CP51 51 5.12 3.84
CP52 52 9.83 2
Spiral Adapter 70 105 0.677
Thermosel DIN spindle 81 3.7 1.29
SSA DIN spindle
for 13R or 13RP chamber 82 3.75 1.29
SSA DIN spindle
for 7R or 7RP chamber 83 12.09 1.29
ULA DIN spindle 85 1.22 1.29
ULA DIN spindle 86 3.65 1.29
ULA DIN spindle 87 12.13 1.29
SPINDLE ENTRY CODE SMC SRC
LVDV-III 0.09373 LV
2.5xLVDV-III 0.2343 4L
5xLVDV-III 0.4686 5L
1/4 RVDV-III 0.25 1R
1/2 RVDV-III 0.5 2R
RVDV-III 1 RV
HADV-III 2 HA
2xHADV-III 4 3A
2.5xHADV-III 5 4A
HBDV-III 8 HB
2xHBDV-III 16 3B
2.5xHBDV-III 20 4B
5xHBDV-III 40 5B
Table D2
MODEL TK
MODEL CODE
ON DV-III SCREEN
Full Scale Viscosity Range [cP] = TK SMC
10,000
RPM
*
*