Installation
Type 8025 / 8035
Install the device on the pipe in such a way that the upstream and downstream distances are respected according to the
design of the pipes, refer to standard EN ISO 5167-1 and "Fig. 8".
flow direction
50 x DN 5 x DN
40 x DN 5 x DN
25 x DN 5 x DN 20 x DN 5 x DN
18 x DN 5 x DN 15 x DN 5 x DN
With control valve Pipe with 2 elbows at 90° in 3
dimensions
Pipe with 2 elbows at 90° Pipe with 1 elbow at 90° or 1
T-piece
With pipe expansion With pipe reduction
Fig. 8: Upstream and downstream distances depending on the design of the pipes
Ensure that the pipe is always filled in the section around the device (see "Fig. 9").
When mounting vertically ensure that the flow direction is in an upward direction (see "Fig. 9").
Horizontal mounting
Correct Incorrect
Vertical mounting
Correct Incorrect
flow direction
Fig. 9: Filling of the pipe
Prevent the formation of air bubbles in the pipe in the section around the device (see "Fig. 10").
Correct
flow direction
Incorrect
Correct Incorrect
Fig. 10: Air bubbles within the pipe
If necessary, use a flow conditioner to improve measurement accuracy.