User guide

46
operating and commissioning
Type 8025 / 8035
9. OPERATING AND COMMISSIONING
9.1. Safety instructions
WARNING
Risk of injury due to nonconforming operating.
Non-conforming operating could lead to injuries and damage the device and its surroundings.
The operators in charge of operating must have read and understood the contents of this manual.
In particular, observe the safety recommendations and intended use.
The device/installation must only be operated by suitably trained staff.
WARNING
Danger due to non-conforming commissioning.
Non-conforming commissioning could lead to injuries and damage the device and its surroundings.
Before commissioning, make sure that the staff in charge have read and fully understood the contents of the manual.
In particular, observe the safety recommendations and intended use.
The device / the installation must only be commissioned by suitably trained staff.
Before commissioning, set the K Factor of the fitting used. See chap. "9.6.3".
9.2. Operating levels of the device
The device has two operating levels: the Process level and the Configuration level.
The Process level makes it possible:
to read the flow rate measured by the device, the value of the current transmitted on the 4-20 mA analogue output, the
values of both the daily and main totalizers;
to reset the daily totalizer;
to access the Configuration level.
The Configuration level has two menus (Parameters, Test) and makes it possible:
to set the device parameters;
to test some device settings.
Tab. 6: Default settings of the device
Function Default value
LANGUAGE English
UNIT of the flow rate L/min, 1 decimal
UNIT of the totalizers L
Number of decimal positions 1
4 mA current output 0.000
20 mA current output 100.0
K FACTOR
51.20
FILTER
5
Value of a pulse (PU)
00.05
Unit of the pulse output
L
Relay 1- 40.0
Relay 1+ 50.0
Relay 1 inverted no
Relay 2- 80.0
Relay 2+ 90.0
Relay 2 inverted no
English