User guide

55
operating and commissioning
Type 8025 / 8035
Sense of operation not inverted Sense of operation inverted
ON
OFF
1-
(2-) (2+)
1+
Contact
Inverted NO
Flow rate
Inverted YES
Flow rate
Contact
ON
OFF
1-
(2-) (2+)
1+
Fig. 62: Hysteresis operating
9.6.7. Configuring the filter of the measured flow rate
This parameter makes it possible to dampen the fluctuations:
of the display;
of the current output.
Ten filters are available.
The following table gives the response time for each filter (10% to 90%):
FILTER Response time FILTER Response time
0 0,15 s 5 6 s
1 0,7 s 6 10 s
2 1,4 s 7 19 s
3 2,5 s 8 33 s
4 3,5 s 9 50 s
FiLtER FiLtER 0
tOtAL
Choose the filter.
Confirm.
FiLtER 9
ENTER
0......9
Fig. 63: Diagram of the "FILTER" parameter of the Parameters menu
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press
ENTER
to
save the settings and go back to the Process level.
9.6.8. Resetting both totalizers
This parameter makes it possible to reset both totalizers.
tOtAL RES NO
END
0......9
Choose to reset the two totalizers or not.
Confirm.
RES yES
ENTER
Fig. 64: Diagram of the "TOTAL" parameter of the Parameters menu