Maintenance and troubleshooting
Type 8025 / 8035
Message
displayed
Possible cause Recommended action
ERROR 7 Memory read error: the user settings
and the totalizer values are lost.
Press the ENTER key to acknowledge the error: the totalizers are
reset.
Send the device back to your supplier.
PWR FAIL
The supply voltage is too low.
The voltage at the device terminals
is lower than 12 V.
The impedance of the current
measurement loop is too high (see
chap. "6.5").
Adjust the supply voltage so that the voltage at the device ter-
minals is between 12 V and 36 V.
PU H LIM The pulse value times the K factor of
the device is > 1000000.
The entered volume for a pulse is too
high.
Enter a lower volume / pulse. See chap. "9.6.5".
Check the K factor value. See chap. "9.6.3".
PU L LIM The pulse value times the K factor of
the device is < 1.
The entered volume for a pulse is too
low.
Enter a higher volume / pulse. See chap. "9.6.5".
Check the K factor value. See chap "9.6.3".