8.5. Details of the Process level
This level is active by default when the device is energized.
12.6 l/s
87654 l
231 l.
0......9
> 2 s
Resetting the daily
totalizer.
Value of the measured flow rate, displayed
in the unit chosen in the "UNIT" parameter of
the Configuration level.
Value of the main totalizer, total volume of
fluid counted by the device since the last
reset.
Value of the daily totalizer (identified by a
dot after the volume units), volume of fluid
counted by the device since the last reset.
0......9
+
Fig. 14: Details of the Process level
8.6. Details of the Configuration level
To access the Configuration level, simultaneously press keys
ENTER
for at least 5 s.
This level comprises the following configurable parameters:
lANGUAGE
Going back to the Process level
and saving the settings.
UNit
K-FACtOR
FiltER
tOtAl
END
0......9
Choosing the display language
Choosing the flow rate unit, the
number of decimals and the unit
the totalizers are displayed in.
Entering the K factor of the fitting
used or have it defined through a
teach-in procedure.
Choosing the degree of attenuation
of the measured flow rate varia-
tions, with effect on the displayed
flow rate.
Resetting both totalizers.
Fig. 15: Diagram of the Configuration level
18
Operating and commissioning
Type 8025/8035 battery powered