User guide

35
Operating and functions
Determine the fitting K factor using a teach-in procedure depending on a volume ("TEACH V.")
The device will use the new K factor as soon as "SAVE YES" is confirmed when leaving the Parameters
menu.
Prepare a tank with a known volume.
Stop the fluid circulation.
confirm "TEACH V.": "FILL END." is displayed.
K-FACtOR K=2.8500
tEACh V.
REtURN
0......9
The device calculates the K factor of the fitting
and displays it.
FiLL END
0000.0 L
VALiD N/Y
tEACh F.
OUtPUt
Charge the pipe to fill the tank.
When the tank is full, confirm "FILL END":
K=2.9000
Enter the volume (value between 0,1 and 9999,9) of fluid
that passed in the circuit. Confirm.
The display shows the K factor of the fitting, may it have been entered
or determined by a teach-in procedure.
Confirm the displayed value.
confirm the K factor determined through teach-in or not.
Fig. 30: Teach-in procedure depending on a volume
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press
ENTER
to save the settings or not and go back to the Process level.
English
Type 8045