User guide

51
Operating and functions
8.6.15. Configuring the electric network frequency
This parameter makes it possible to configure the electric network frequency so that the device can filter the inter-
fering signals of the power supply.
Adjust this parameter even if the device operates with a DC voltage.
FREqUENC. F = 50hz
CUt-OFF
0......9
Choose the electric network frequency.
Confirm
F = 60hz
Fig. 51: Diagram of the "FREQUENC." parameter of the Parameters menu
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press
ENTER
to save the settings or not and go back to the Process level.
8.6.16. Parameterizing the cut-off flow rate
This parameter makes it possible to set the flow rate value to 0 if the measured value is less than the set cut-off
value:
• the display then shows a flow rate = 0 (a dot is displayed after the flow rate units).
• the outputs and the totalizers react as if the actual flow rate was = 0.
CUt-OFF CO=0.000
BACKLit
Enter the measured flow rate value below which the device considers that the flow rate
is = 0.
To disable the function, set CO=0.
Confirm
Fig. 52: Diagram of the "CUT-OFF" parameter of the Parameters menu
If you do not want to adjust another parameter, go to the "END" parameter of the Parameters menu and press
ENTER
to save the settings or not and go back to the Process level.
English
Type 8045