User Manual

Notes
12
MC19-1.20
5.6 Running a Program
The current program can be started from the home display by pressing RUN/HOLD. Or by
pressing page once, then scroll once (twice for P10 and P25), then press up or
down
5.7 To pause (hold) a program
Press RUN/HOLD
or
Press page until Program Status Reset appears
Press scroll until the cursor moves to Reset
Press up or down to select Hold
RUN/HLD will be displayed
5.8 To stop and reset a program
Press and hold RUN/HOLD
or
Press page until Program Status Reset appears
Press scroll until the cursor moves to Reset
Press up or down to select Reset
5.9 To run a different program (P10 and P25)
To select a program
Press page until Program Status Reset appears
Press scroll then up or down to select a program number
To run the program
Press RUN/HOLD
or
Press scroll until the cursor moves to Reset
Press up or down to select Run
Run will be displayed
5.10 Program Status
While a program is running the home display shows 3 values:
Top: Measured temperature
Centre: Program setpoint (PSP)
Bottom: Power output percentage
To see more detail:
Press page once to go to the program status page. The top and centre values from the
home page remain on the display. The lower half of the screen now shows:
Current program number (P10 and P25 only)
Current segment number
Time remaining for that segment
Further presses of the scroll key while a program is running will reveal additional
information:
Status. This can be changed to Hold, Reset or Run by pressing up or down .