If the signal check still does not illuminate,
change the transmitting cable. Hold the
signal check over the transmitting coil and
speak into the microphone. If the signal
check light illuminates with incoming
sound, then the original transmitting cable is
faulty.
5. If the signal check still does not illuminate, plug
the lapel microphone into the speech processor
external input socket - thereby disconnecting
the headset microphone. Align and hold the
signal check over the transmitting coil. If the
signal check light illuminates with incoming
sound, then the problem is in the headset
microphone and you should contact your
implant center.