User manual

64 SPrint User Manual
CARING FOR THE SYSTEM
Headset function check
1. Turn on the speech processor and ensure the
coil is directly over the cochlear implant.
2. Set the microphone sensitivity to 10. Speak into
the headset microphone.
If the segment meter responds to speech, then
the headset and its connection to the speech
processor are functioning correctly.
3. If the segment meter does not respond to
speech, replace the headset cable with a spare
cable and check the segment meter in response
to speech.
4. If the segment meter still does not respond to
speech, replace the transmitting cable with a
spare cable and check the meter in response to
speech.
5. If the segment meter still does not respond to
speech, connect the lapel microphone to the
speech processor external input socket and
check the meter in response to speech.
If you still have a problem, contact your implant
center.
N02210F ISS4 DO2.qxd 24/4/02 1:12 PM Page 64