Operation
25
Fault Displays
FieldMaxII-TO is capable of detecting internal and user-
induced faults. When a fault is detected, the letters
“Er”—followed by a numeric fault code (see
Table 1)—appear on the display. You can dismiss a fault
code by pressing any button, or by correcting the cause
of the fault.
Table 1. Fault Codes
ERROR CONDITION FAULT CODE
Unrecognized sensor 1
Sensor communication failure 2
Sensor error 3
Sensor error 4
Sensor/firmware version mismatch (sensor
format version exceeds capability of the instru-
ment firmware—firmware upgrade needed)
5
Sample rate fault 6
Hardware fault (detectable hardware error) 20
Bad zero
a
40
Data overflow (result of an arithmetic opera-
tion that is greater than can be held in the allo-
cated storage)
41
Wrong type of sensor is attached to the instru-
ment
42
a. For more information about the bad zero fault code, refer to “Zero” on page 33.