36
result computed. If the Restart mode is Auto, a new batch
is immediately restarted. The error that caused the
contaminated batch is displayed.
Measurement
Display and
Range
Selection
The display update rate for numeric measurement is
three times per second.
Measurement range is selected in decade steps. Range
selection (shown in Table 2 on page 37) is dependent on
the sensor type and characteristics, as well as user-deter-
mined measurement settings.
FieldMaxII-TO uses the “3’s” Rule—a display format-
ting rule in which the display value is not allowed to
exceed 3, 30, or 300, depending on where the decimal
point falls, with the decimal point located in a fixed posi-
tion, as determined by the current range. Typically if a
reading exceeds the “3’s” limit, the instrument is over-
ranged or, in the case of Auto Ranging, the instrument
will automatically range up.