Item Brochure

Specifications subject to change without notice.
Microstream, Microcap, and FilterLine are registered trademarks and Molecular Correlation Spectroscopy and Uni-junc-
tion are trademarks of Oridion Medical 1987 Ltd. The capnography component of this product is covered by one or more
of the following US patents: 6,428,483; 6,997,880 ; 6,437,316; 7,488,229; 7,726,954 and their foreign equivalents.
COVIDIEN, COVIDIEN with logo, Covidien logo and
positive results for life
are U.S. and internationally registered
trademarks of Covidien AG. Other brands are trademarks of a Covidien company. ™* and ®* are trademarks of their
respective owners. ©2013 Covidien. 12-PM-0253(1)
 G A
B, CO

--
..
Specifications and Accessories
CAPNOGRAPHY
Display Range: 0-99 mmHg
0-13.0 Vol %
0-13.2 kPa
Sampling flow rate 50 ml/ min
Warm-up time 30 seconds typical
Accuracy
0-38 mmHg ± 2 mmHg
39-99 mmHg ± 5% of reading
Respiratory rate 0-150 breaths/min
Modes Adult, neonatal
Alarms Adjustable alarm limits for etCO
2
, FiCO
2
, and respiratory rate
Display Graphic LCD display (128 x 64 dots) with LED backlight
(contrast and backlight control); Bright, seven-segment
LED for numeric display of etCO
2
and respiratory rate; Two
color alarm bar
Nurse Call RS232 interface between monitor and remote systems.
and Bernoulli MSM Nurse Call System can be easily wired via stereo phone jack.
Interface Microcap
only - one parameter (v3.15 and higher):
Compatibility: Output to Philips CMS 2001, V24, V26 via VueLink A01
and A05. Output to Spacelabs Ultraview
®*
, PCMS
™*
MONITORS
Microcap
North America/U.S. CS04178
Outside North America/Non-U.S. CS04179
ACCESSORIES
Product Name Part Number
Carrying Case CS04321
Monitor Protective Boot CS07779
Clamp/Mounting Bracket CS03743
Rechargeable Battery Pack (replacement for original battery pack) CS03736
Medical Grade AC Adapter (replacement for original AC adapter)
US CS04934
EU CS08610
12 Volt cable CS08505
D to A Converter (typically used to interface to sleep labs) CS07143
Communication Adapter Kit CS04284
MSM Interface Kit (Bernoulli, Oxinet) CS08644
Nurse Call Interface Kit CS08642
A complete set of interface and connectivity options
Nurse Call Interface provides remote alarm connectivity to existing call system
Bernoulli MSM wireless technology platform enables data transfer to central monitoring systems
Interfaces with Philips (via VueLink) and Spacelabs (via Flexport®*) monitoring systems
STANDARDS AND REGULATIONS
EN 60601-1, EN 60601-1-2, EN ISO 21647 (capnography), UL60601-1, CSA
C22.2 No.601.1-M90, applicable parts of EN 1789 (when used with protective
boot), CE mark and 510(k) cleared
FILTERLINE
®
ETCO
2
SAMPLING LINES
FilterLine
®
CO
2
sampling line solutions are available for all patients, whether
intubated or non-intubated, in any clinical setting. The unique Microstream
patient interfaces reliably sample oral and nasal breathing, even in the presence
of supplemental O
2
.