Product Manual

IN TRANSPORT, IN THE HOSPITAL, AT HOME
Sometimes your patients need short-term
continuous monitoring — wherever they are. That’s
why we developed the Nellcor
exible SpO
2
sensor.
Our new rubber pulse oximetry sensor:
Is reusable
Is designed for short-term (less than six hours)
continuous monitoring for the majority of
your patients
Comes in two sizes — large and small —
with a >20 kilogram weight range, one of the
broadest available
Features the pulse oximetry you trust-
Nellcor
SpO
2
with OxiMax
technology
BECAUSE
FLEXIBILITY
MAT TERS.
Nellcor
Flexible SpO
2
Sensor
Our sensor is forward and backward compatible with
Nellcor
TM
-compatible pulse oximetry technology.
And it’s compliant with the latest standards. Like
those for rotary and xed-wing aircraft. Our sensor
has been tested to meet:
Air worthiness vibration testing for both xed and
rotary winged aircraft
Electromagnetic compatibility (radiated emissions)
levels in accordance with standard DO-160G
It meets extreme environment standards.
We’ve tested our sensor from:
0 to 40 C operating temperature
15 to 95 percent non-condensing relative
humidity (RH)
620 to 1060 hPa atmospheric pressure
All in compliance with IEC 60601-1-12:2014 and
an ingress protection rating of IP33. (IP33 protects
against solid particulates >2.5 mm and spraying
water up to 60 degrees from vertical with no harmful
eects.)
Our sensor will operate for 20 minutes at
temperatures down to minus 20 C and up to 50 C.
When not in use, you can store our sensor in
conditions of:
From -40 C up to 70 C
15 to 95 percent non-condensing RH

Summary of content (2 pages)