User Manual

SYMPTOM PROBABLE CAUSE CORRECTIONS
Hold Error Pump in HOLDING mode for
more than 10 minutes.
Press
}
CONTINUE to return to HOLDING screen, or
press and hold the Power button to turn off. Several options
are available from the HOLDING screen, such as RUN or
ADJUST SETTINGS.
Flow Error Occlusion in pump-patient line.
(downstream occlusion)
Check for and correct the occlusion in the line between pump
and patient. Replace Pump Set if error cannot be resolved.
Feed Error Empty bag or occlusion in
bag-pump line. (upstream
occlusion)
Check for empty bag and refill, or check for and correct
the occlusion in the line between the pump and feed bag.
Replace Pump Set if error cannot be resolved.
Flow Error
Use > 24 Hrs
Pump Set in use beyond
recommended length of time.
Informational warning message blinks on upper left of screen.
It is recommended that the Pump Set be replaced.
Pump Set
Dislodged
Pump Set not properly loaded
to pump.
Check black ring retainer (MISTIC) and valve; reload Pump
Set and restart pump to continue. Replace Pump Set if error
cannot be resolved.
Battery Low Battery charge too low. Immediately plug pump into A/C outlet to recharge battery
and to provide power for pump.
Rotor Error Pump Set tubing not properly
loaded on rotor, or some
other unusual rotor operating
condition.
Check that the tubing is loaded on the rotor and that it is not
damaged, torn, etc. Reload tubing or replace Pump Set.
System Error General error caused by many
factors. See manual for error list.
Power down pump and attempt restart to clear error. Call
customer service with screen code if error cannot be resolved.
Menu Selections
Programming & Menu
Selection Buttons
Status LEDs
Red=Error
Yellow=Hold/Pause
Green=Normal Operation
Power Button
15 Hampshire Street
Mansfield, MA
02048
1-800-962-9888
508-261-8000
www.covidien.com
COVIDIEN, COVIDIEN with logo, Covidien logo,
positive results for life
and Kangaroo
are U.S. and/or internationally registered trademarks of Covidien AG.
©2010 Covidien.
H6206-10M-0610
Listed below are some of the probable causes of alarm conditions and their corrections.
Kangaroo
Joey Enteral Feeding Pump
Troubleshooting Guide