Loading & Rinsing of Feeding Pump
Do not use feeding sets for longer than 24 hours.
Proper loading & unloading of a feeding set
Proper rinsing of the feeding set
1) Feed & flush valve assembly, 2) Feed only valve
assembly, 3) Distal end of feeding set, 4) Pump
1) Grasp thumb tab on valve and insert into left
pocket, ensure valve is fully seated. Thumb tab
should align with slot in blue door.
2) Grasp black ring retainer, gently wrap tubing
around rotor, and insert retainer directly into
right pocket. Do not overstretch tubing.
1) Disconnect feeding set from patient before rins-
ing. The feeding set must remain loaded during
the rinsing procedure.
2) CAUTION- if set is removed from the pump,
it may damage the set and prevent use and will
negativley effect the system accuracy.
1
2
3
4
3) To properly unload a pump set, power the
pump off and wait 10 seconds before removing
the feeding set.
3) Empty contents of feed bag.
4) Fill the feed bag with water to approximatley
half of the total capacity. Properly seal the bag
and shake for 5 to 10 seconds.