User guide

OPERATION Protocol Plus Owners Manual
22 Version 2
Copyright © 2010 by Despatch Industries.
All rights reserved. No part of the contents of this manual may be reproduced, copied or transmitted in any form or by any
means including graphic, electronic, or mechanical methods or photocopying, recording, or information storage and
retrieval systems without the written permission of Despatch Industries, unless for purchaser's personal use.
O
Control can be configured to automatically activate Manual,
Timer or Profile Mode when power is applied (power switch ON).
Refer to Section 5.5.
5.2. Manual Mode
1. Press Select until Manual displays.
O
Press Run at any time in the procedure to activate Manual Mode.
2. Press Menu to display the Process Temperature Setpoint (Setpt). Change Setpoint
(Setpt) by pressing or .
O
If the SPChange on the Enable page in Setup Mode has been set
to DISABLED, it must be changed to ENABLED before any
changes to the process temperature and high limit setpoints can
be made.
3. Press
Menu a second time to display current high limit setpoint (HLSP). The value
can be adjusted pressing
or . If Band displays, the High Limit band feature is
activated (see Control page, Section 5.6.7) and the high limit cannot be adjusted.
4. Optional feature: Press
Menu a third time to display Event1. Press to turn the
event ON or to turn the even OFF. Repeat for all events which are enabled (up to
four events).
5. To start Manual Mode, press Run.
The display changes from Stop to Run. Return to Stopped Mode by pressing Stop. While
in operation, the process setpoint can be adjusted by using or to change the value
while the mode is running. Press Menu to display the High Limit Setpoint (HLSP)
setting.
O
Changes to High Limit Setpoint or Event Output Configuration
must be accomplished from the Stop Mode.