User's Manual

12
A-DV54
EN
ADVAnceD menu controls
Advanced Menu List
Display Example Mode Description
All
This is the default screen when the IntelliPAP is
turned on. The actual mask pressure and heater
setting are shown. The displayed pressure may vary
slightly while breathing. If the heater system is not
connected, the heat setting will be blank.
AutoAdjust
Mode
This is a prescription setting which limits the
maximum pressure at which the IntelliPAP will
operate if in AutoAdjust mode. Your home care
provider programs this setting for you during device
setup.
AutoAdjust
Mode
This is a prescription setting which sets the
minimum pressure at which the IntelliPAP will
operate in AutoAdjust mode. Your home care
provider programs this setting for you during device
setup.
CPAP
Mode
This is a prescription setting which controls the
operating pressure.
All
Selectable delay time setting. Your provider has set
the delay time. If the unlocked symbol is shown,
you are able to adjust this setting. If the locked
symbol is shown, this setting can only be adjusted
by your provider.
CPAP
Mode
This is a prescription setting which controls the
operating pressure during the delay time.
All
The standard tubing length supplied with the device
is 6 feet (1.8 meters). However, you can adjust your
IntelliPAP to accommodate an air supply tubing
length of 10 feet (3.0 meters). Replacement tubing
can be obtained through your home care provider
(part numbers listed at the end of this manual).
All
Pressing the down key enters the Enable menu.
Refer to next section for Enable Menu details.
All
This setting changes the contrast for the LCD
display. This only needs to be adjusted if you find it
difficult to see the display from your viewing angle.
All
The total number of hours (in 1/10 hour
increments) that the IntelliPAP has been providing
therapy.
All
The total number of hours (in 1/10 hour
increments) that the IntelliPAP has been in
operation.
Compliance Meter
123456.7 hours
Hourmeter
123456.7 hours
LCD Contrast Adj
50 %
Enable Menu
Press
P: 5.0 Auto
Heat:Off
Upper Limit
15.0 cmH20
Lower Limit
5.0 cmH20
Set Pressure
10.0 cmH20
Delay Time
45 minutes
Tubing Length
1.8 m (6 ft)
Delay Pressure
5.0 cmH20