Specifications

14 DC-530 Operation Manual
displayed in the NEW SETTING field.
5. Once you have entered your change, you have the following options:
press the + (Plus) key to temporarily store the change and move to the next specification, then press
the * PRINT key to permanently record the change and store it and leave the SPEC setting mode.
Press the TARE key to exit without saving the change.
6. To change another SPEC code before exiting, repeat Steps 3 and 4.
If you power off the scale while the saving process in going on, you can clear the entire memory of the
scale! Wait until the scale has returned to the Menu Screen before performing any other functions.
7. After returning to the Normal Weighing Mode, power the scale down completely by using the ON/OFF
rocker switch on the bottom of the scale to complete the registering of the changes.
SPEC No Bit 3 Bit 2 Bit 1 Bit 0
20 Minimum Display (Uni Bloc)
00: 1 [DEFAULT]
01: 2
10: 5
11: 10
Display Resolution (Uni Bloc)
(when internal count span is 500,000)
00: 1/10,000
01: 1/5,000 [DEFAULT]
10: 1/2,500
11: 1/12,500
(when internal count span is 1,000,000)
00: 1/100,000
01: 1/50,000 [DEFAULT]
10: 1/25,000
11: 1/20,000
21 Internal Count Span (Uni
Bloc)
0: 500,000
1: 1,000,000 [DEFAULT]
Weight Decimal Point Position (Uni Bloc)
(when internal count span is 500,000)
000: 00000
001: 0000.0 [DEFAULT]
010: 000.00
011: 00.000
100: 0.0000
Weight Decimal Point Position (Uni Bloc)
(when internal count span is 1,000,000)
000: 000000
001: 00000.0 [DEFAULT]
010: 0000.00
011: 000.000
100: 00.0000
101: 0.00000
22 not currently in use
Table 3-2. DC-530 Weight and Measurement Specifications