104 Bio-inert Quaternary Pump User Manual
8 Test Functions and Calibration
Leak Rate Test
Running the Test
Running the test from the Agilent Lab Advisor
1 Select the Leak Rate Test from the Test Selection menu.
2 Start the test and follow the instructions.
Evaluating the Results
Results of the leak rate test are the leak rates measured for pistons 1 and 2 as
described for the test principle. If any of the leak rates exceeds 3 µL/min, the
test will fail.
Parts required p/n Description
01080-83202 Blank nut
Make sure to release the pressure by slowly opening the purge valve when the test has
finished.