User manual

Bio-inert Quaternary Pump User Manual 11
Introduction
1
Overview of the Hydraulic Path
Overview of the Hydraulic Path
The bio-inert quaternary pump is based on a two-channel, dual-piston
in-series design which comprises all essential functions that a solvent delivery
system has to fulfill. Metering of solvent and delivery to the high-pressure side
are performed by one pump assembly which can generate pressure up to
600 bar.
Degassing of the solvents is done in a built-in vacuum degasser. Solvent
compositions are generated on the low-pressure side by a high-speed
proportioning valve (MCGV).
The pump assembly includes a pump head with an active inlet valve and an
outlet valve. A damping unit is connected between the two piston chambers. A
purge valve including a PTFE frit is fitted at the pump outlet for convenient
priming of the pump head.
The pump includes an active seal wash function, which cleans pistons and
seals for the use of buffer solutions.