Bio-inert Quaternary Pump User Manual 123
Maintenance
9
Removing the Pump Head Assembly
1
Remove the front cover by pressing the clip fasteners on
both sides of the cover.
2 Disconnect the active inlet valve cable.
3 Using a 1/4 inch wrench remove the outlet capillary. 4 Remove the waste tubing and disconnect the solvent
tubing from the inlet valve.
5 Remove the capillary at the bottom of the pump head.