User manual

Bio-inert Quaternary Pump User Manual 125
Maintenance
9
Seal Wear-in Procedure
Seal Wear-in Procedure
1 Place a bottle with 100 mL of isopropanol in the solvent cabinet and put a
tubing (including bottle head assembly) into the bottle.
2 If an active inlet valve is installed, screw the PEEK adapter 1/4-28 to 10-32
(p/n 0100-1847) to the AIV and connect the inlet tube from the bottle head
directly to it.
3 Connect the Restriction capillary (p/n 5022-2159) to the purge valve.
Connect its other end to a waste container.
4 Open the purge valve and purge the system for 5 minutes with isopropanol
at a flow rate of 2 mL/min.
5 Close the purge valve and set the flow to a rate adequate to achieve a
pressure of 350 bar. Pump 15 min at this pressure to wear in the seals. The
pressure can be monitored using your instrument control software or tool.
6 Turn OFF the pump, slowly open the purge valve to release the pressure
from the system, disconnect the restriction capillary and reinstall the bottle
with the solvent for your application.
7 Rinse your system with the solvent used for your next application.
8 Replace the purge valve frit, see “Exchanging the Purge Valve Frit or the
Purge Valve” on page 119.
Parts required p/n Description
0100-1847 Adapter AIV to solvent inlet tubes
5022-2159 Restriction capillary
Seal damage
This procedure is required for Standard seals (pack of 2) (p/n 5063-6589), but it will
damage the PE seals.
The stainless steel restriction capillary is acceptable for a temporary test within a bio-inert
system. In case of concerns, a PEEK capillary (e.g. PEEK capillary, 150 cm length, 0.13 mm
i.d. (p/n 0890-1915)) and PEEK fittings (e.g. PEEK Fittings 10/PK (p/n 5063-6591)) may be
used instead.