Bio-inert Quaternary Pump User Manual 131
Maintenance
9
Exchanging the Multi-Channel Gradient Valve (MCGV)
Exchanging the Multi-Channel Gradient Valve (MCGV)
Tools required p/n Description
8710-0899 Screwdriver, Pozidriv #1
Parts required p/n Description
G5611-67701 Multi-Channel Gradient Valve (MCGV) (Bio-inert)
Preparations Switch off pump at the main power switch
Remove the front cover
Use a solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding leakages.
The life time of the multi-channel gradient valve can be increased by regularly flushing the
valve, especially when using buffers. If using buffers, flush all channels of the valve with
water to prevent precipitation of the buffer, otherwise salt crystals could drop into an
unused channel and form plugs that may cause leaks of that channel. Such leaks will
interfere with the general performance of the valve. When using buffers in combination
with organic solvents in the Agilent 1260 Infinity Bio-inert Quaternary Pump it is
recommended to connect the aequous solutions/buffers to one of the bottom ports (A and
D) and the organic solvent to one of the upper gradient valve ports. It is best to have the
organic channel directly above the buffer channel (e.g., A - buffer, B - organic solvent).