Bio-inert Quaternary Pump User Manual 17
Introduction
1
Bio-inert Materials
Bio-inert Materials
For the Agilent 1260 Infinity Bio-inert LC system, Agilent Technologies uses
highest quality materials in the flow path (also referred to as wetted parts),
which are widely accepted by life scientists, as they are known for optimum
inertness towards biological samples and ensure best compatibility to common
samples and solvents over a wide pH range. Explicitly, the complete flow path
is free of stainless steel and free of other alloys containing metals like iron,
nickel, cobalt, chromium, molybdenum or copper, which can interfere with
biological samples. The flow downstream of the sample introduction does not
contain metals at all.
Tab le 1 Bio-inert Materials
Module Materials
Agilent 1260 Infinity Bio-inert Quaternary Pump
(G5611A)
Titanium, gold, platinum-iridium,
zirconium oxide, ruby, PTFE, PEEK
Agilent 1260 Infinity Bio-inert High-Performance Autosampler
(G5667A)
Upstream of sample-introduction:
Titanium, gold, PTFE, PEEK,
zirconium oxide
Downstream of
sample-introduction:
PEEK, zirconium oxide
Agilent 1260 Infinity Bio-inert Manual Injector
(G5628A)
PEEK, zirconium oxide
Agilent 1260 Infinity Bio-inert Analytical Fraction Collector
(G5664A)
PEEK, zirconium oxide, PTFE