Bio-inert Quaternary Pump User Manual 179
Hardware Information
12
Setting the 8-bit Configuration Switch (without On-board) LAN
Special Settings
The special settings are required for specific actions (normally in a service
case).
Boot-Resident
Firmware update procedures may require this mode in case of firmware
loading errors (main firmware part).
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident mode. It is not operable as a
module. It only uses basic functions of the operating system for example, for
communication. In this mode the main firmware can be loaded (using update
utilities).
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Tab le 17 Boot Resident Settings (without on-board LAN)
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
No LAN TEST/BOOT11001000
Loss of data
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are diagnosis and repair log books which will not be erased.
Save your methods and data before executing a forced cold start.
Tab le 18 Forced Cold Start Settings (without on-board LAN)
Mode Select SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
No LAN TEST/BOOT11001001