80 Bio-inert Quaternary Pump User Manual
7 Error Information
Module Error Messages
Module Error Messages
These errors are pump specific.
Restart Without Cover
The module was restarted with the top cover and foam open.
The sensor on the main board detects when the top foam is in place. If the
module is restarted with the foam removed, the module switches off within
30 s, and the error message is generated.
Solvent Zero Counter
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings
in the data system. If the volume level in the bottle falls below the specified
value the error message appears when the feature is configured accordingly.
Probable cause Suggested actions
1 The module started with the top cover and
foam removed.
Please contact your Agilent service
representative.
2 Rear of the module is exposed to strong
direct sunlight.
Ensure that the rear of module is not directly
exposed to strong sunlight.
Probable cause Suggested actions
1 Volume in bottle below specified volume.
Refill bottles and reset solvent counters.
2 Incorrect setting.
Make sure the limits are set correctly.