User manual

90 Bio-inert Quaternary Pump User Manual
7 Error Information
Module Error Messages
Index Limit
The time required by the piston to reach the encoder index position was too
short (pump).
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the index position is reached too fast, the error
message is generated.
Index Adjustment
The encoder index position in the module is out of adjustment.
During initialization, the first piston is moved to the mechanical stop. After
reaching the mechanical stop, the piston reverses direction until the encoder
index position is reached. If the time to reach the index position is too long,
the error message is generated.
Probable cause Suggested actions
1 Irregular or sticking drive movement.
Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
2 Defective pump drive assembly.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Irregular or sticking drive movement.
Remove the pump head, and examine the seals,
pistons, and internal components for signs of
wear, contamination or damage. Exchange
components as required.
2 Defective pump drive assembly.
Please contact your Agilent service
representative.